A certificate of analysis, usually shortened to COA, is the document that connects a specific batch of a research compound to measured laboratory results. If you buy research peptides, it is the single most useful piece of paper a supplier can give you. This guide walks through each section of a typical peptide COA, explains what the numbers mean, and lists the warning signs that a report may not be worth trusting.
- Lot first. The lot number on the report must match the one on your vial, or the report describes different material.
- Purity (HPLC) is the main peak’s share of peak area, read at 214 to 220 nm. Look for the chromatogram, not just a number.
- Identity (mass spec) gives a measured mass in daltons that should sit within the stated tolerance of the theoretical mass.
- Net peptide content is the share of the powder that is peptide. The rest is mostly water and counter-ions, so it is below 100%.
- Accountability. A named lab, dated signatures and a way to confirm the report with the lab.
See an annotated sample COAExplain your own COA
Find the 15 red flags hidden in two sample reports. It takes about two minutes.
What a certificate of analysis is
A COA is a report issued by a laboratory, or by a supplier’s quality department, that lists the tests performed on one particular lot of material and the results of each test. The key phrase is one particular lot. A COA for lot A says nothing about lot B, even if both carry the same product name. That is why the lot or batch number is the first thing to check.
Section 1: Identification
A complete COA names the compound, the lot or batch number, the quantity per vial, the manufacturing or testing date and the laboratory. Check four things:
- The compound name and sequence or structure match what you ordered.
- The lot number on the report matches the lot number printed on your vial.
- The date of testing is plausible for the batch you were shipped.
- The testing laboratory is identified by name, not just by a logo.
If the lot number does not match, the document describes some other material, and you should ask the supplier for the correct report before relying on it.
Section 2: Appearance
Most lyophilized peptides are described as a white to off-white powder or cake. Appearance is a simple visual test, but it still matters. Unexpected colour, a collapsed or melted cake, or visible particles can point to moisture exposure, contamination or degradation.
Section 3: Purity by HPLC
Purity is almost always measured by reversed-phase high-performance liquid chromatography, or RP-HPLC. The sample is dissolved and pushed through a column. Different molecules leave the column at different times, and a detector, usually reading ultraviolet light around 214 to 220 nanometres where peptide bonds absorb, draws each one as a peak. Purity is the area of the main peak divided by the total area of all peaks.
- A result such as 98.5 percent means the main peak makes up 98.5 percent of the detected signal.
- The chromatogram itself should be attached or at least described. A single large peak with a clean baseline is what you want to see.
- Many small peaks near the main one often mean closely related impurities from synthesis, such as truncated or deleted sequences.
Purity by HPLC is relative to what the detector can see. It does not tell you about water, salts or compounds that do not absorb at the measured wavelength.
Section 4: Identity by mass spectrometry
HPLC can show a clean peak for the wrong molecule. Mass spectrometry closes that gap. The instrument measures the mass-to-charge ratio of ions and, from that, the molecular weight of the compound. The measured mass should match the theoretical mass for the stated sequence within the tolerance of the instrument. Electrospray ionisation and MALDI-TOF are the common methods. A COA that gives purity but no identity result is incomplete.
Section 5: Net peptide content and water
Lyophilized powder is not 100 percent peptide. It also contains water and counter-ions, often trifluoroacetate or acetate. Net peptide content, sometimes measured by amino acid analysis or elemental nitrogen, tells you what fraction of the powder is actual peptide. Water content is commonly measured by Karl Fischer titration. These values matter when you calculate how much peptide is really in a vial for quantitative work.
Section 6: Optional tests
Some suppliers also report endotoxin by the LAL assay for cell-based work, residual solvents, microbial limits or counter-ion content. Their presence is a good sign of a more complete quality program, but their absence is not unusual for standard research-grade material.
Red flags in a COA
- No lot number, or a lot number that does not match the vial.
- No laboratory name, signature or date.
- A purity figure with no chromatogram and no mass result.
- The same chromatogram appearing in reports for different compounds or lots.
- Round numbers such as 99.9 percent on every product.
- Low-resolution or cropped images where axes and peak labels cannot be read.
Practice on a sample report
The fastest way to learn a COA is to read one. Tap through our interactive sample COA, test yourself in the spot the fake COA challenge, or print the sample below. When you have a real report in hand, upload it to the COA explainer for a line-by-line walkthrough.
Free to print and share for education. All data is illustrative and comes from a fictional laboratory.
Frequently asked questions
Is a higher purity number always better? Not necessarily. Method, column and wavelength all affect the figure, and a very high number without identity confirmation tells you little.
Who should run the test? An independent laboratory removes the conflict of interest that comes with a supplier testing its own products, and accreditation to ISO/IEC 17025 is a useful marker of quality systems.
How do I find the report for my lot? Search the lot number in our COA library, or request a quote if you need documentation for a specific order. For more on the two main tests, read our guide to HPLC purity versus mass spectrometry.