Most certificates of analysis you will see are genuine. The problems tend to come from three places: reports that have been edited, reports that belong to a different lot than the vial in your hand, and real numbers that are easy to misread. This challenge trains your eye for all three.
Below are two sample peptide COAs with 15 red flags hidden between them. Tap anything that looks wrong. When you finish, grab the free printable checklist and use it on every report you receive.
The single best check is to call or email the testing laboratory yourself, using the contact details on the lab’s own website rather than the ones printed on the report. Give them the sample ID and ask them to confirm the results. A lab can confirm its own report; it cannot confirm one it never issued.
The challenge
Round 1 hides 7 red flags and round 2 hides 8. The easy ones are in the layout. The hard ones hide in the numbers: read the units, the limits and whether the figures agree with each other.
123 Sample Way, Suite 100
Anytown, USA 00000
www.example.com · info@example.com
(555) 010-0199
| Client | Example Customer |
|---|---|
| Accession # | 2611-0457 |
| Search code | EXC-2611-0457 |
| Received | 11/03/2026 |
| Tested | 11/04/2026 |
| Lot | BPC-2611-01 |
| Condition | Intact, ambient |
| Revision | 0 (original issue) |
Sample as receivedMethods: reversed-phase HPLC with UV detection; identity by electrospray mass spectrometry (LC-MS); water by Karl Fischer; residual solvents by headspace GC; endotoxin by LAL. References: USP <85> (bacterial endotoxins), USP <921> (water determination), ICH Q3C (residual solvents).
| Test | Method | Method ID | Specification | Result | Disposition |
|---|---|---|---|---|---|
| Appearance | Visual | QC-VIS-001 | White to off-white lyophilized powder | Conforms | Pass |
| Identity | ESI-MS | QC-MS-007 | 1418.70 ± 0.50 Da (monoisotopic) | 1418.71 Da | Pass |
| Purity | RP-HPLC, UV 214 nm | QC-HPLC-014 | ≥ 98.0% (area) | 98.6% | Pass |
| Net peptide content | Amino acid analysis | QC-AAA-003 | Report | 80.4% w/w | — |
| Water | Karl Fischer | QC-KF-002 | ≤ 8.0% w/w | 5.2% w/w | Pass |
| Counter-ion (TFA) | Ion chromatography | QC-IC-005 | Report | 11.8% w/w | — |
| Residual acetonitrile | Headspace GC | QC-GC-009 | ≤ 410 ppm | 120 ppm | Pass |
| Bacterial endotoxin | LAL, kinetic chromogenic | QC-LAL-004 | Report | < 0.5 EU/mg | — |
Swipe the table sideways to see every column.
| Peak | RT (min) | Area % |
|---|---|---|
| 1 | 10.90 | 0.30 |
| 2 | 11.85 | 0.80 |
| 3 | 12.40 | 98.60 |
| 4 | 12.95 | 0.20 |
| 5 | 13.70 | 0.10 |
| Ion | m/z | Charge |
|---|---|---|
| [M+3H]3+ | 473.91 | 3+ |
| [M+2H]2+ | 710.36 | 2+ |
| [M+H]+ | 1419.71 | 1+ |
Date: 2026.11.04 16:42:10 -05’00’
Date: 2026.11.05 09:18:52 -05’00’
Notes: This certificate of analysis reflects only the sample indicated on this form, as received by the laboratory in good condition. This report may not be reproduced except in full. For research use only.
ISO/IEC 17025 accredited testing laboratory
www.example.com/verify
Search code EXC-2611-0457
123 Sample Way, Suite 100
Anytown, USA 00000
www.example.com · info@example.com
(555) 010-0199
| Client | Example Customer |
|---|---|
| Accession # | 2611-0457 |
| Search code | EXC-2611-0457 |
| Received | 11/03/2026 |
| Tested | 10/28/2026 |
| Lot | BPC-2609-07 |
| Condition | Intact, ambient |
| Revision | 0 (original issue) |
Sample as receivedMethods: reversed-phase HPLC with UV detection; identity by electrospray mass spectrometry (LC-MS); water by Karl Fischer; residual solvents by headspace GC; endotoxin by LAL. References: USP <85> (bacterial endotoxins), USP <921> (water determination), ICH Q3C (residual solvents).
| Test | Method | Method ID | Specification | Result | Disposition |
|---|---|---|---|---|---|
| Appearance | Visual | QC-VIS-001 | White to off-white lyophilized powder | Conforms | Pass |
| Identity | ESI-MS | QC-MS-007 | 1418.70 ± 0.50 Da (monoisotopic) | 1419.56 Da | Pass |
| Purity | — | — | — | 99.9% | Pass |
| Net peptide content | Amino acid analysis | QC-AAA-003 | Report | 95.0% w/w | — |
| Water | Karl Fischer | QC-KF-002 | ≤ 8.0% w/w | 5.2% w/w | Pass |
| Counter-ion (TFA) | Ion chromatography | QC-IC-005 | Report | 11.8% w/w | — |
| Residual acetonitrile | Headspace GC | QC-GC-009 | ≤ 410 ppm | 120 ppm | Pass |
| Bacterial endotoxin | LAL, kinetic chromogenic | QC-LAL-004 | Report | < 0.5 ppm | — |
Swipe the table sideways to see every column.
| Peak | RT (min) | Area % |
|---|---|---|
| 1 | 10.90 | 0.02 |
| 2 | 11.85 | 0.06 |
| 3 | 12.40 | 99.90 |
| 4 | 12.95 | 0.01 |
| 5 | 13.70 | 0.01 |
| Ion | m/z | Charge |
|---|---|---|
| [M+3H]3+ | 473.91 | 3+ |
| [M+2H]2+ | 710.36 | 2+ |
| [M+H]+ | 1419.71 | 1+ |
Notes: This certificate of analysis reflects only the sample indicated on this form, as received by the laboratory in good condition. This report may not be reproduced except in full. For research use only.
ISO/IEC 17025 accredited testing laboratory
www.example.com/verify
Search code EXC-2611-0457
Certified · Verified · Trusted
| Client | Example Customer |
|---|---|
| Accession # | 2611-0457 |
| Search code | — |
| Received | 11/03/2026 |
| Tested | 11/04/2026 |
| Lot | BPC-2611-01 |
| Condition | Intact, ambient |
| Revision | 0 (original issue) |
Sample as receivedMethods: reversed-phase HPLC with UV detection; identity by electrospray mass spectrometry (LC-MS); water by Karl Fischer; residual solvents by headspace GC; endotoxin by LAL. References: USP <85> (bacterial endotoxins), USP <921> (water determination), ICH Q3C (residual solvents).
| Test | Method | Method ID | Specification | Result | Disposition |
|---|---|---|---|---|---|
| Appearance | Visual | QC-VIS-001 | White to off-white lyophilized powder | Conforms | Pass |
| Identity | ESI-MS | QC-MS-007 | 1418.70 ± 0.50 Da (monoisotopic) | 710.36 Da | Pass |
| Purity | RP-HPLC, UV 280 nm | QC-HPLC-014 | ≥ 98.0% (area) | 99.2% | Pass |
| Net peptide content | Amino acid analysis | QC-AAA-003 | Report | 85.6% w/w | — |
| Water | Karl Fischer | QC-KF-002 | ≤ 8.0% w/w | 0.0% w/w | Pass |
| Counter-ion (TFA) | Ion chromatography | QC-IC-005 | Report | 11.8% w/w | — |
| Residual acetonitrile | Headspace GC | QC-GC-009 | ≤ 5,000 ppm | 3,900 ppm | Pass |
| Bacterial endotoxin | LAL, kinetic chromogenic | QC-LAL-004 | Report | < 0.5 EU/mg | — |
Swipe the table sideways to see every column.
| Peak | RT (min) | Area % |
|---|---|---|
| 1 | 10.90 | 0.17 |
| 2 | 11.85 | 0.46 |
| 3 | 12.40 | 99.20 |
| 4 | 12.95 | 0.11 |
| 5 | 13.70 | 0.06 |
| Ion | m/z | Charge |
|---|---|---|
| [M+3H]3+ | 473.91 | 3+ |
| [M+2H]2+ | 710.36 | 2+ |
| [M+H]+ | 1419.71 | 1+ |
Date: 2026.11.04 16:42:10 -05’00’
Date: 2026.11.05 09:18:52 -05’00’
Notes: This certificate of analysis reflects only the sample indicated on this form, as received by the laboratory in good condition. This report may not be reproduced except in full. For research use only.
The laboratory, lots and results on these sample reports are fictional and for illustration only.
The most common edit: a changed company name
In practice, the most common change to a COA is not invented data. It is the company name. A reseller who buys from a manufacturer will often replace the client name on the manufacturer’s report with its own, because it does not want to reveal its supplier. To do that, the original PDF usually gets converted into an editable file or an image, edited and re-exported.
Often the results are left exactly as the lab reported them, and the motive is to protect a business relationship rather than to deceive. Even so, the document is no longer the original. Once a report has been edited, you cannot tell from the page alone whether only the name changed. That is why two checks matter more than anything printed on the report:
- Ask for the original PDF. Many accredited labs sign their PDFs digitally. A PDF reader’s signature panel shows who signed the file and whether it has been changed since. A screenshot, a photo or a re-exported file carries no signature at all.
- Confirm the sample ID with the lab. The lab will tell you whether the results on your copy match its records, even if it cannot tell you who the original client was.
A trustworthy seller should either provide the original report or be open that it has been relabeled, and should not object to you confirming the results with the lab.
What is normal and should not worry you
Real reports also scare people for the wrong reasons. These results look alarming at first glance but are completely normal:
- Residual solvents. Almost any purified material contains a trace of the solvents used to make it. Seeing acetonitrile or another solvent listed is expected. It is a threshold question: the amount should sit below the limit, such as 410 ppm for acetonitrile under ICH Q3C.
- Net peptide content well below 100%. Lyophilized peptides hold water and counter-ions such as TFA, so a powder that is 98% pure by HPLC can still be only 75 to 85% peptide by weight.
- A few percent water. Lyophilized peptides absorb moisture from the air. Some water is expected; exactly 0.0% is the suspicious result.
- “< LOQ” and “ND”. These mean below the method’s quantitation or detection limit. They are usually good news, not missing data.
- mg/g next to percent. Many labs report both. 1% equals 10 mg/g, so 98.6 mg/g and 9.86% are the same result written two ways.
For a full walk-through of every unit on a report, see Peptide COA results and units explained, which includes free calculators.
Round 1: the 7 red flags explained
Lot number does not match the vial
The report covers lot BPC-2609-07, but the vial says BPC-2611-01. Every result on the page describes different material. Always match the lot number character by character.
Tested before it was received
The test date falls six days before the receipt date. Impossible dates usually mean a report was edited or recycled from an older one.
Average mass passed against a monoisotopic specification
1419.56 Da is BPC-157’s average mass. The specification is the monoisotopic mass, 1418.70 ± 0.50 Da, so the result should have failed. It was marked Pass anyway.
Purity with no method
A round 99.9% with no method, wavelength or specification cannot be checked. Real HPLC results name the method and are rarely round.
Weight percentages over 100%
95.0% peptide, 5.2% water and 11.8% TFA add up to 112%. A powder cannot be more than 100% of itself.
Endotoxin in the wrong unit
Endotoxin is reported in endotoxin units (EU/mg), not ppm. A unit that does not fit the test is a sign the report was not produced by people who run it.
No sign-off
No analyst, no reviewer and no dates. Accredited labs have a named analyst and a second reviewer sign every report.
Round 2: the 8 red flags explained
A screenshot instead of the original PDF
The report arrived as an image. Converting a PDF to an image is usually how reports get edited, and images carry no digital signature. Ask for the original PDF.
Slogans instead of lab details
“Certified, verified, trusted” replaces the address, contact details and accreditation. Without them you cannot contact the lab or check its credentials.
The wrong sequence
BPC-157 is GEPPPGKPADDAGLV. The report reads GEPPPGKPAEDAGLV, with glutamic acid where aspartic acid should be. The identity check was either done against the wrong structure or the text was copied from another document.
An m/z value reported as the mass
710.36 is the m/z of the doubly charged ion, not the molecular mass in daltons. Passing it against a 1418.70 Da specification shows the spectrum was misread.
Purity measured at 280 nm
BPC-157 has no tryptophan or tyrosine, so it barely absorbs at 280 nm, and neither do most of its impurities. Peptide purity belongs at 214 to 220 nm.
Exactly 0.0% water
Lyophilized peptides always hold some moisture. A Karl Fischer result of exactly zero suggests the number was typed rather than measured.
A loosened specification
Acetonitrile’s ICH Q3C limit is 410 ppm. Raising the specification to 5,000 ppm lets a high 3,900 ppm result pass. Check the limit, not just the word Pass.
Missing pages
The footer says page 1 of 3, but there is only one page. Missing pages often hold the chromatogram, spectra or failed tests.
How to check that a COA is genuine: 6 steps
- Contact the lab directly. Use the phone number or email on the lab’s own website, quote the sample ID, and ask them to confirm the results.
- Get the original PDF. Open it in a PDF reader and check the signature panel for the signer and whether the file has been modified.
- Match the lot. The lot number on the report must match your vial exactly.
- Check the accreditation. Look the lab up on its accreditation body’s website (in the United States, bodies such as ANAB and A2LA) and confirm the tests on your report are within its ISO/IEC 17025 scope.
- Read the numbers properly. Check units, limits and whether weight percentages add up to about 100%. Our units guide and calculators help, or upload the report to our COA explainer to have the math checked for you.
- Make sure you have everything. Every page, the chromatogram and the mass spectrum, not just a summary line.
Get the printable checklist
Keep a one-page version of these checks next to your bench. Enter your email and the PDF downloads right away.
One printable page with the checks from this guide: confirming the report with the lab, file and document checks, the numbers that matter, and what is normal.
By signing up you agree to receive email updates from G2 MEDS. Products are for in-vitro laboratory research only.
Training a team or teaching a class? Print the practice pack: both flawed reports, ready to mark up, with an answer key for all 15 red flags.
Free to print and share for education. All data is illustrative and comes from a fictional laboratory.
Share or embed the challenge
Teaching a team or writing about lab quality? The challenge is free to embed on any website for educational use.
Free to use for education. Paste this code into any page.
Frequently asked questions
What is the best way to tell if a COA is fake?
Contact the testing laboratory directly, using the contact details on the lab’s own website rather than those printed on the report, and ask it to confirm the sample ID and results. A lab can confirm a report it issued and will tell you if it never issued it.
Is a PDF COA more trustworthy than a screenshot?
Generally, yes. Many accredited labs digitally sign their PDFs, and a PDF reader shows whether the file has been changed since it was signed. Screenshots and photos carry no signature and are easy to edit. Even with a PDF, confirming the results with the lab is the strongest check.
Why would a seller change the company name on a COA?
Resellers often replace the client name with their own so they do not reveal the manufacturer they buy from. The results are frequently left unchanged, but the document is no longer the original, so ask for the original file or confirm the sample ID with the lab.
Is it bad if a COA shows residual solvents?
No. Almost any purified material contains a trace of the solvents used to make it. What matters is whether the amount is below the limit, for example 410 ppm for acetonitrile under the ICH Q3C guideline.
What does ISO/IEC 17025 accreditation mean?
It means an independent accreditation body has assessed the laboratory’s quality system and technical competence for a defined list of tests. Check that the tests on your report fall within the lab’s accredited scope.
Are the reports in this challenge real?
No. The laboratory, lots and results are fictional and were created for education. The red flags are based on problems that appear on unreliable reports.
Related reading
- Peptide molecular weight calculator
- Peptide COA results and units explained (with calculators)
- How to read a certificate of analysis for research peptides
- Third-party testing: what independent labs actually measure
- HPLC purity vs. mass spec
- How to choose a research peptide supplier: a checklist
This guide is educational and describes laboratory documentation only. It is not use, dosing or administration guidance. G2 MEDS products are sold strictly for in-vitro laboratory research and are not for human or veterinary use.